A combination of magnetic permeability detection with nanometer-scaled superparamagnetic tracer and its application for one-step detection of human urinary albumin in undiluted urine

M Lu, Filiz Ibraimi, D Kriz, Kirstin Kriz

Research output: Contribution to journalArticlepeer-review

Abstract

A rapid (6.5 min) and simple one-step magnetic immunoassay (MIA) has been developed for analysis of human urinary albumin in near patient settings. Polyclonal rabbit anti-human albumin was used as a capture antibody and monoclonal mouse anti-human albumin as a detection antibody in a two-site immunometric assay requiring no additional washing procedures. The polyclonal anti-human albumin was conjugated to silica microparticles (solid phase) and the monoclonal antibody to dextran-coated nanoscaled superparamagnetic particles (tracer). Quantification of human albumin in undiluted urine was performed by adding 2 mu L urine to a measuring vial containing solid-phase, superpararnagnetic tracer and reaction buffer and then inverting the vial by hand for 20 s. The measuring vial was allowed to stand for 6 min prior to detection, in order for the solid-phase sediment to form at the bottom of the vial. Lastly, the measuring vial was placed into a magnetic permeability detector, which measured the enrichment of superparamagnetic tracer in the sediment due to complex formation with human albumin. Total analysis time was 6.5 min. A linear response was obtained for 0-400 mg/L albumin with a detection limit of 5 mg/L. The total coefficient of variation (CV) was 11% calculated from four consecutive runs on a urine sample containing 11.1 mg/L human albumin during 3 consecutive days. Human urinary albumin analysis was performed on 149 patient samples using the MIA technique and the obtained results showed good correlation with the hospital immunoturbidimetric reference method (y = 1.004x + 4.01, R-2 = 0.978, N= 149) and a commercially available point of care albumin analysis provided by HemoCue Inc. (y = 0.98x + 5.8, R-2 = 0.833, N= 90). (c) 2005 Elsevier B.V. All rights reserved.
Original languageEnglish
Pages (from-to)2248-2254
JournalBiosensors & Bioelectronics
Volume21
Issue number12
DOIs
Publication statusPublished - 2006

Subject classification (UKÄ)

  • Biochemistry and Molecular Biology

Free keywords

  • tracer
  • superparamagnetic
  • anti-human albumin-conjugated
  • albumin-conjugated silica particles
  • anti-human
  • magnetic immunoassay (MIA)
  • human urinary albumin