An experimental study of the dose response of polymer gel dosimeters imaged with x-ray computed tomography

J V Trapp, Sven Bäck, M Lepage, G Michael, C Baldock

Research output: Contribution to journalArticlepeer-review

Abstract

Changes in the linear attenuation coefficient of polymer gel dosimeters post-irradiation enable the imaging of dose distributions by x-ray computed tomography (CT). Various compositions of polymer gel dosimeters manufactured from acrylamide (AA), and N,N'-methylene-bis-acrylamide (BIS) comonomers and gelatin or agarose gelling agents were investigated. This work shows that increasing the comonomer concentration increases the CT-dose sensitivity of the polymer gel dosimeter. This can be further increased by replacing gelatin with agarose. Varying the gelatin concentration however does not significantly change the CT-dose sensitivity. Among the compositions studied, dose resolution (D(delta)95%) was found to be optimal for polymer gel dosimeters comprising 5% gelatin, 3% AA, 3% BIS and 89% water.
Original languageEnglish
Pages (from-to)2939-2951
JournalPhysics in Medicine and Biology
Volume46
Issue number11
DOIs
Publication statusPublished - 2001