Assessment of precision, concordance, specificity, and sensitivity of islet cell antibody measurement in 41 assays

E. Bonifacio, C. Boitard, H. Gleichmann, M. A. Shattock, J. L. Molenaar, G. F. Bottazzo, S. Assa, A. Arnaiz-Villena, J. Barbosa, C. Betterle, E. Beutner, G. Bright, H. Chapel, M. Codina, R. Dawkins, E. Deitsch, U. Di Mario, G. Eisenbarth, R. Elliot, R. Gomis de BarbaraT. Hanafusa, L. Harrison, K. Helmke, C. Howard, P. In t. Veld, D. Kawathara, T. Kobayashi, M. Landin, Å Lernmark, N. Maclaren, T. Mandrup-Poulsen, R. Manna, A. Miettinen, J. Palmer, P. Panczel, J. Peter, K. Pirich, L. Quenette, G. Reeves, K. Reinauer, W. Scherbaum, L. Scott-Morgan, D. Vergani, B. Vialettes

Research output: Contribution to journalArticlepeer-review

Abstract

Forty-one assays were analysed at the 3rd International Workshop on the standardisation of islet cell antibodies. Analysis of precision demonstrated assays consistently detecting blind duplicates within one doubling dilution and capable of discriminating one doubling dilution differences in islet cell antibody concentration. Some assays, however, reported duplicates discrepantly by more than seven doubling dilutions, and consequently could not distinguish even large quantities of islet cell antibodies. Precision was best in assays from laboratories which had participated in all three Standardisation Workshops and was not dependent upon methodology. The use of the Juvenile, Diabetes Foundation reference islet cell antibody standard and standard curves reduced the scatter of results, and was best amongst assays with better precision. Twenty-seven assays reported all ten blood donor sera as negative. However, 14 assays did not, and specificity (negativity in health) was <50% in three assays. Low specificity was strongly associated with poor precision. The detection limit of assays ranged from <5 to 50 JDF units and was partially dependent upon methodology. Assays incorporating extended incubation had the lowest detection limits without a decrease in the specificity of the ten blood donor sera. Precise quantification is fundamental for the standardisation and comparability of islet cell antibodies. Precise quantitative assays have been identified and reference standards and common units established.

Original languageEnglish
Pages (from-to)731-736
JournalDiabetologia
Volume33
Issue number12
DOIs
Publication statusPublished - 1990 Dec 1
Externally publishedYes

Subject classification (UKÄ)

  • Endocrinology and Diabetes

Free keywords

  • Islet cell antibody
  • Juvenile Diabetes Foundation units
  • quality control
  • standards
  • Type 1 (insulin-dependent) diabetes

Fingerprint

Dive into the research topics of 'Assessment of precision, concordance, specificity, and sensitivity of islet cell antibody measurement in 41 assays'. Together they form a unique fingerprint.

Cite this