N2 - Abstract Objectives. To determine whether a steerable catheter with electrogram guidance (CS-assist group) could facilitate access to the coronary sinus (CS) during cardiac resynchronization therapy (CRT) implantation. Design. Consecutive patients who underwent CRT implantation were recruited prospectively into the CS-assist group (n = 81) and compared with those using conventional techniques without an electrogram guidance (conventional group, n = 101). Results. The CS cannulation success rate was clearly greater in the CS-assist group (100%) than that in the conventional group (95%, p < 0.05), with significantly shorter mean procedure time (52.6 ± 20.6 min vs. 73.2 ± 40.9 min, p < 0.01) and fluoroscopy time (3.6 ± 3.2 min vs. 14.2 ± 20.4 min, p < 0.01). In the five CS cannulation failure cases, mean procedure time (144.0 ± 37.0 min) and fluoroscopy time (57.8 ± 24.8 min) were significantly longer than those in the other patients (61.2 ± 32.3 and 8.2 ± 13.6 min, respectively, n = 177, both p < 0.01). Conclusions. Using the steerable catheter with real-time electrogram guidance, location of and access to the CS is more rapid and successful, which may improve the success of the CRT implantation and may give significant time savings.
AB - Abstract Objectives. To determine whether a steerable catheter with electrogram guidance (CS-assist group) could facilitate access to the coronary sinus (CS) during cardiac resynchronization therapy (CRT) implantation. Design. Consecutive patients who underwent CRT implantation were recruited prospectively into the CS-assist group (n = 81) and compared with those using conventional techniques without an electrogram guidance (conventional group, n = 101). Results. The CS cannulation success rate was clearly greater in the CS-assist group (100%) than that in the conventional group (95%, p < 0.05), with significantly shorter mean procedure time (52.6 ± 20.6 min vs. 73.2 ± 40.9 min, p < 0.01) and fluoroscopy time (3.6 ± 3.2 min vs. 14.2 ± 20.4 min, p < 0.01). In the five CS cannulation failure cases, mean procedure time (144.0 ± 37.0 min) and fluoroscopy time (57.8 ± 24.8 min) were significantly longer than those in the other patients (61.2 ± 32.3 and 8.2 ± 13.6 min, respectively, n = 177, both p < 0.01). Conclusions. Using the steerable catheter with real-time electrogram guidance, location of and access to the CS is more rapid and successful, which may improve the success of the CRT implantation and may give significant time savings.
UR - https://www.scopus.com/pages/publications/84893332829
U2 - 10.3109/14017431.2013.875623
DO - 10.3109/14017431.2013.875623
M3 - Article
SN - 1651-2006
VL - 48
SP - 41
EP - 46