Abstract
Due to the involvement of calpains and calpastatin in the pathology of several diseases, it is important to create an analytical tool for their accurate and sensitive measurements. In this work, an amperometric immunoassay (AIA) was developed and optimized for quantification of two key proteins from the calpain proteolytic system. The detection is based on the measurement of a steady-state current on screen printed electrodes due to the reduction of the enzymatic product resulting from specific immunoassays, where the measured current is proportional to the concentration of the target analyte. The optimized assays with dynamic ranges from 0.09 to 6 ng/mL for calpastatin and 0.04 to 5 mu g/mL for calpain. offer simplicity, sensitivity and possibility for miniaturization. (C) 2011 Elsevier B.V. All rights reserved.
Original language | English |
---|---|
Pages (from-to) | 248-253 |
Journal | Sensors and Actuators B: Chemical |
Volume | 152 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2011 |
Subject classification (UKÄ)
- Biological Sciences
- Industrial Biotechnology
Free keywords
- Amperometric immunoassay
- Calpastatin
- Calpain
- Pathology