Skip to main navigation Skip to search Skip to main content

Evaluation of digital breast tomosynthesis systems

Gustav Hellgren, Thahn Tra Pham, Anders Tingberg, Magnus Dustler

Research output: Chapter in Book/Report/Conference proceedingPaper in conference proceedingpeer-review

Abstract

In this study, two digital breast tomosynthesis (DBT) systems were evaluated: Siemens Mammomat Inspiration TOMO (Siemens Healthineers, Erlangen, Germany) and GE Senographe Pristina (GE, Buc, France). Along with differences such as angular range and detectors type, the systems use different reconstruction algorithms. One was available for the GE system, based on iterative reconstruction (IR). Two algorithms were available for the Siemens system: TOMO_STANDARD, using filtered back projection (FBP) and EMPIRE, FBP with statistically based artifact reduction. Two commercially available DBT phantoms (CIRS model 020 & 021), with heterogeneous and homogenous background respectively, were used to calculate signal-difference-to-noise-ratio (SDNR) in key structures for varying phantom thickness (30, 45 & 70 mm) and average glandular dose (AGD). Key phantom structures include calcifications and lesion masses of different sizes. Results show a positive correlation between SDNR and AGD except for the EMPIRE algorithm where there was a negative SDNR/AGD trend for one of the microcalcification specks in the heterogeneous phantom. The highest overall SDNR was acquired using the EMPIRE algorithm. Both systems are well within the recommended dose limits but could increase their dose levels in order to achieve higher SDNR. This indicates that there may be room for dose optimization in DBT systems used in screening programs, confirming the importance of continuous evaluation and optimization.

Original languageEnglish
Title of host publicationMedical Imaging 2020
Subtitle of host publicationPhysics of Medical Imaging
Editors/authorsGuang-Hong Chen, Hilde Bosmans
PublisherSPIE
ISBN (Electronic)9781510633919
DOIs
Publication statusPublished - 2020
EventMedical Imaging 2020: Physics of Medical Imaging - Houston, United States
Duration: 2020 Feb 162020 Feb 19

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume11312
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2020: Physics of Medical Imaging
Country/TerritoryUnited States
CityHouston
Period2020/02/162020/02/19

Subject classification (UKÄ)

  • Radiology and Medical Imaging

Free keywords

  • Breast Imaging
  • Breast Screening
  • DBT
  • Digital breast tomosynthesis
  • Dose optimization

Fingerprint