FORWARD-SCATTERED RADIATION FROM THE COMPRESSION PADDLE SHOULD BE CONSIDERED IN GLANDULAR DOSE ESTIMATIONS.

Bengt Hemdal

Research output: Contribution to journalArticlepeer-review

6 Citations (SciVal)

Abstract

From major protocols on dosimetry in mammography, there is no doubt that the incident air kerma should be evaluated without backscattered radiation to the dosemeter. However, forward-scattered radiation from the compression paddle is neglected. The aim of this work was to analyse the contribution of forward-scattered radiation for typical air kerma measurements. Measurements of forward-scatter were performed with a plane-parallel ionisation chamber on four mammography units. The forward-scatter contribution to the air kerma was 2-10 % and increased with the compression paddle thickness, but also with the half-value layer value. For incident air kerma in mammography, it can be as important to consider forward-scattered as backscattered radiation. If an ionisation chamber is used, the compression paddle should be in contact with the chamber; otherwise the air kerma and absorbed dose will be underestimated. If a dosemeter based on semiconductors with much less sensitivity to scattered radiation is used, it is suggested that a forward-scatter factor (FSF) is applied. Based on the results of this work, FSF=1.06 will lead to a maximum error of ∼4 %.
Original languageEnglish
Pages (from-to)196-201
JournalRadiation Protection Dosimetry
Volume147
DOIs
Publication statusPublished - 2011

Subject classification (UKÄ)

  • Radiology, Nuclear Medicine and Medical Imaging

Fingerprint

Dive into the research topics of 'FORWARD-SCATTERED RADIATION FROM THE COMPRESSION PADDLE SHOULD BE CONSIDERED IN GLANDULAR DOSE ESTIMATIONS.'. Together they form a unique fingerprint.

Cite this