High-resolution diffusion imaging using phase-corrected segmented echo-planar imaging

Sara Brockstedt, J R Moore, C Thomsen, Stig Holtås, Freddy Ståhlberg

Research output: Contribution to journalArticlepeer-review

26 Citations (SciVal)

Abstract

Diffusion magnetic resonance imaging (MRI) was performed with a high-resolution segmented echo-planar imaging technique, which provided images with substantially less susceptibility artifacts than images obtained with single-shot echo-planar imaging (EPI). Diffusion imaging performed with any multishot pulse sequence is inherently sensitive to motion artifacts and in order to reduce motion artifacts, the presented method utilizes navigator echo phase corrections, performed after a one-dimensional Fourier transform along the frequency-encoding direction. Navigator echo phases were fitted to a straight line prior to phase correction to avoid errors from internal motion. In vivo imaging was performed using electro cardiographic (ECG) triggering. Apparent diffusion coefficient (ADC) maps were calculated on a pixel-by-pixel basis using up to seven diffusion sensitivities, ranging from b = 0 to 1129 x 10(6) s/m(2).