Abstract
BACKGROUND: There are limited data guiding the optimal inflation time for coronary drug-coated balloons. The aim of this study was to assess the clinical efficacy of an inflation time of ≥30 seconds compared with <30 seconds.
METHODS: We conducted an observational nationwide analysis using the SCAAR (Swedish Coronary Angiography and Angioplasty Registry) to identify all patients undergoing percutaneous coronary intervention with drug-coated balloons and their inflation times. Outcomes were assessed at 1 year using Kaplan-Meier estimates and adjusted Cox regression.
RESULTS: The <30-second group comprised 412 patients, and the ≥30-second group 10 826 patients. Patients in the <30-second group presented more often with lesions in proximal segments. After a 1-year follow-up, ≥30 seconds was associated with lower rates of target segment revascularization compared with <30 seconds (3.2% versus 6.4%, adjusted hazard ratio [HR], 0.47 [95% CI, 0.32-0.69], P≤0.001) and target vessel myocardial infarction (1.8% versus 5.3%, adjusted HR, 0.32 [95% CI, 0.21-0.50], P≤0.001). No difference was observed for all-cause mortality.
CONCLUSIONS: In this analysis, including 11 238 patients undergoing PCI with drug-coated balloons, ≥30 second inflation time was associated with improved outcomes compared with a shorter inflation time.
| Original language | English |
|---|---|
| Article number | e047354 |
| Journal | Journal of the American Heart Association |
| Volume | 15 |
| Issue number | 15 |
| DOIs | |
| Publication status | Published - 2026 Aug 4 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Subject classification (UKÄ)
- Cardiology and Cardiovascular Disease
Free keywords
- Humans
- Female
- Time Factors
- Registries
- Angioplasty, Balloon, Coronary/instrumentation
- Treatment Outcome
- Male
- Aged
- Coronary Artery Disease/therapy
- Coronary Angiography
- Coated Materials, Biocompatible
- Middle Aged
- Sweden/epidemiology
- Cardiac Catheters
- Cardiovascular Agents/administration & dosage
- Risk Factors
- Percutaneous Coronary Intervention/instrumentation
- Myocardial Infarction
Fingerprint
Dive into the research topics of 'Impact of Drug-Coated Balloon Inflation Time on Outcome After Percutaneous Coronary Intervention: An Observational Study From SCAAR'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver