Abstract
The relative size and relaxation of the invisible pool of bound spins (T1b) underlying magnetization transfer (MT) was quantified in eight subjects in vivo at 1.5 T from progressive saturation experiments using repetitive MT pulses. The evolution of the binary spin–bath was sampled by increasing the repetition period from 8 to 200 ms. Single-shot echo-planar images at TE=50ms were evaluated in the central white matter. Three models were fitted: the general solution, and with constraints of equal relaxation and T1b=1s for the invisible pool. The general solution of unconstrained T1b provided a significantly better fit, indicating fast-to-intermediate exchange. The bound pool fraction was 17±4%, the relaxation times T1f=1.6±0.2s for free water and T1b=171±22ms for the bound pool. The constrained models did not differ from each other, since here T1b was similar to the observed T1 of 1.1±0.1s. They underestimate the bound pool fraction and its relaxation. Thus, the standard assumption of continuous-wave MT models may underestimate the relaxation via the bound pool by more than a factor of five.
Original language | English |
---|---|
Pages (from-to) | N529-N540 |
Journal | Physics in Medicine and Biology |
Volume | 54 |
Issue number | 23 |
DOIs | |
Publication status | Published - 2009 |