Abstract
A new, fast, and automated sample pretreatment technique for determination of lipophilic organic compounds in aqueous samples has been developed and applied to the determination of polychlorinated biphenyls (PCBs) in environmental river water. It is based on miniaturized microporous membrane liquid-liquid extraction coupled on-line to gas chromatography (GC) with electron capture detection. The heart of the system that simultaneously connects the sample pretreatment step to the final GC analysis has been named the extracting syringe (ESy). The ESy carries a miniaturized membrane extraction card attached to an electrically and mechanically designed installment and is mounted directly over a GC injector for fully automated injection of the extract. A method was developed to extract 10 PCB congeners from 1-mL water samples (after addition of 40% acetonitrile) with an extraction time of 10 min. The optimized methodology showed good linearity (in the dynamic concentration range of 5 ng L-1-1 mug L-1), enrichment factors of 33-40 times, repeatable extractions (RSD 2-5%, n = 4), a
Original language | English |
---|---|
Pages (from-to) | 1928-1934 |
Journal | Analytical Chemistry |
Volume | 76 |
Issue number | 7 |
DOIs | |
Publication status | Published - 2004 |
Bibliographical note
The information about affiliations in this record was updated in December 2015.The record was previously connected to the following departments: Analytical Chemistry (S/LTH) (011001004)
Subject classification (UKÄ)
- Analytical Chemistry