Abstract
Pregnancy increases the risk of pulmonary embolism. Computed tomography pulmonary angiography (CTPA) is used for diagnosis. CT generates ionising radiation, and thus, abdominal shielding may be used. This phantom study investigated the effects of patient shielding and scan length reduction on the fetal and maternal ionising radiation dose from CTPA. The absorbed dose to the fetus was measured using thermoluminescent dosemeters. Estimated effective doses to the pregnant patient were based on the dose-length products. Shielding increased both the effective dose to the patient by 47% and the mean absorbed dose to the fetus (0.10 vs. 0.12 mGy; p < 0.001) compared with unshielded standard CTPA, as it affected the automatic exposure control. Shielded short CTPA marginally lowered only the mean fetal absorbed dose (0.03 vs. 0.02 mGy; p = 0.018). Shortening the scan reduced the fetal absorbed dose most effectively by 70% (0.10 vs. 0.03 mGy; p = 0.006), compared with the standard unshielded scan. Shielding modestly reduces fetal radiation dose but may compromise automatic exposure control, possibly increasing the maternal and fetal radiation dose. Shortening the scan is beneficial, assuming anatomical coverage is secured.
Original language | English |
---|---|
Pages (from-to) | 458-465 |
Number of pages | 8 |
Journal | Radiation Protection Dosimetry |
Volume | 189 |
Issue number | 4 |
DOIs | |
Publication status | Published - 2020 Jul 24 |