A high-repetition rate scheme for synchrotron-based picosecond laser pumpx-ray probe experiments on chemical and biological systems in solution

Research output: Contribution to journalArticle

Abstract

We present the extension of time-resolved optical pump/x-ray absorption spectroscopy (XAS) probe experiments towards data collection at MHz repetition rates. The use of a high-power picosecond laser operating at an integer fraction of the repetition rate of the storage ring allows exploitation of up to two orders of magnitude more x-ray photons than in previous schemes based on the use of kHz lasers. Consequently, we demonstrate an order of magnitude increase in the signal-to-noise of time-resolved XAS of molecular systems in solution. This makes it possible to investigate highly dilute samples at concentrations approaching physiological conditions for biological systems. The simplicity and compactness of the scheme allows for straightforward implementation at any synchrotron beamline and for a wide range of x-ray probe techniques, such as time-resolved diffraction or x-ray emission studies.

Details

Authors
  • Frederico A. Lima
  • Christopher J. Milne
  • Dimali C.V. Amarasinghe
  • Mercedes Hannelore Rittmann-Frank
  • Renske M.Van Der Veen
  • Marco Reinhard
  • Susanne Karlsson
  • Steven L. Johnson
  • Daniel Grolimund
  • Camelia Borca
  • Thomas Huthwelker
  • Markus Janousch
  • Frank Van Mourik
  • Rafael Abela
  • Majed Chergui
External organisations
  • California Institute of Technology
  • Paul Scherrer Institute
  • Swiss Federal Institute of Technology
  • Lund University
Research areas and keywords

Subject classification (UKÄ) – MANDATORY

  • Atom and Molecular Physics and Optics
Original languageEnglish
Article number063111
JournalReview of Scientific Instruments
Volume82
Issue number6
Publication statusPublished - 2011 Jun
Publication categoryResearch
Peer-reviewedYes
Externally publishedYes