Fast Timing with Plastic Scintillators for In-beam Heavy-ion Spectroscopy

Research output: Contribution to journalArticle

Abstract

The design, R&D, and testing of a new plastic-scintillator detector for Time-of-Flight measurements with relativistic heavy-ion beams are presented. A design approach using 32 independent precise timing measurements of the same physical event is followed. This is different from the conventional scheme, which aims at two or four high-precision measurements. A circular, 27 cm in diameter, BC-420 plastic-scintillator sheet is read-out by 32 photomultiplier tubes in order to achieve an intrinsic detector resolution on the order of 10 Ps root mean square. (C) 2011 Elsevier B.V. All rights reserved.

Details

Authors
Organisations
Research areas and keywords

Subject classification (UKÄ) – MANDATORY

  • Accelerator Physics and Instrumentation

Keywords

  • Plastic scintillator, Heavy ion, Time-of-Flight, Fast timing, Time, resolution
Original languageEnglish
Pages (from-to)354-360
JournalNuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment
Volume654
Issue number1
Publication statusPublished - 2011
Publication categoryResearch
Peer-reviewedYes

Related research output

Hoischen, R., 2011, Department of Physics, Lund University. 138 p.

Research output: ThesisDoctoral Thesis (compilation)

View all (1)

Related projects

Robert Hoischen, Dirk Rudolph, Pavel Golubev & J Gerl

2006/10/012011/03/31

Project: Dissertation

View all (1)