Initial Phantom Validation of Minute Roughness Measurement Using Phase Tracking for Arterial Wall Diagnosis Non-Invasively In Vivo

Research output: Contribution to journalArticle

Abstract

To detect minute roughness, we utilized the phase change that occurs in a radio frequency echo from the rough surface of an object during its lateral motion. The new method was optimized and validated using saw-tooth-shaped silicone phantoms sized from 13 to 33 mu m; results were compared to those obtained using a confocal laser scanning microscope.

Details

Authors
Organisations
Research areas and keywords

Subject classification (UKÄ) – MANDATORY

  • Medical Engineering
Original languageEnglish
Pages (from-to)853-857
JournalIEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Volume58
Issue number4
Publication statusPublished - 2011
Publication categoryResearch