Is a hollow centralizer necessary when using a polished, tapered, cemented femoral stem? A randomized, controlled RSA study of 60 hips with 10 years of follow-up

Research output: Contribution to journalArticle

Standard

Harvard

APA

CBE

MLA

Vancouver

Author

RIS

TY - JOUR

T1 - Is a hollow centralizer necessary when using a polished, tapered, cemented femoral stem?

T2 - A randomized, controlled RSA study of 60 hips with 10 years of follow-up

AU - Weber, Erik

AU - Olsson, Christer

AU - Kesteris, Uldis

AU - Flivik, Gunnar

PY - 2017/7/4

Y1 - 2017/7/4

N2 - Background and purpose — A tapered, polished and collarless stem is normally equipped with a hollow centralizer to prevent the stem from becoming end-bearing in the cement as the stem subsides. In a randomized clinical trial, we evaluated such a stem (MS-30), which was initially introduced with a solid centralizer but was later recommended to be fitted with a hollow centralizer. We hypothesized that while the stem would sink more, it would become rotationally stable and have less retroversion with a hollow centralizer than with a solid centralizer. Patients and methods — We randomized 60 patients with primary hip arthritis to receive either a hollow centralizer or a solid centralizer with the stem. The effect was evaluated over a 10-year follow-up period with repeated RSA examinations, conventional radiographs, and clinical follow-ups using the WOMAC and SF-12 questionnaires. Results — At 10-year follow-up, the group with hollow centralizers had subsided more than the group with solid centralizers (1.99 mm (hollow) as opposed to 0.57 mm (solid); p < 0.001). However, rotation was similar at 10-year follow-up (mean retroversion 1.34° (hollow) and 1.30° (solid)). Both groups showed excellent 10-year results, with similar clinical outcome, and none of the stems were radiographically loose or had been revised. Interpretation — As expected, there was more subsidence in the group with hollow centralizers, and with similar magnitude to that reported in earlier RSA studies on conceptually similar prostheses. Interestingly, there was no difference in the rotational behavior of the prostheses. This stem type appears to have a design that, regardless of the type of centralizer and the possibility of subsidence, withstands the rotational forces it is subjected to very well. This study does not support the need for a hollow centralizer for these types of stems.

AB - Background and purpose — A tapered, polished and collarless stem is normally equipped with a hollow centralizer to prevent the stem from becoming end-bearing in the cement as the stem subsides. In a randomized clinical trial, we evaluated such a stem (MS-30), which was initially introduced with a solid centralizer but was later recommended to be fitted with a hollow centralizer. We hypothesized that while the stem would sink more, it would become rotationally stable and have less retroversion with a hollow centralizer than with a solid centralizer. Patients and methods — We randomized 60 patients with primary hip arthritis to receive either a hollow centralizer or a solid centralizer with the stem. The effect was evaluated over a 10-year follow-up period with repeated RSA examinations, conventional radiographs, and clinical follow-ups using the WOMAC and SF-12 questionnaires. Results — At 10-year follow-up, the group with hollow centralizers had subsided more than the group with solid centralizers (1.99 mm (hollow) as opposed to 0.57 mm (solid); p < 0.001). However, rotation was similar at 10-year follow-up (mean retroversion 1.34° (hollow) and 1.30° (solid)). Both groups showed excellent 10-year results, with similar clinical outcome, and none of the stems were radiographically loose or had been revised. Interpretation — As expected, there was more subsidence in the group with hollow centralizers, and with similar magnitude to that reported in earlier RSA studies on conceptually similar prostheses. Interestingly, there was no difference in the rotational behavior of the prostheses. This stem type appears to have a design that, regardless of the type of centralizer and the possibility of subsidence, withstands the rotational forces it is subjected to very well. This study does not support the need for a hollow centralizer for these types of stems.

U2 - 10.1080/17453674.2017.1315553

DO - 10.1080/17453674.2017.1315553

M3 - Article

VL - 88

SP - 377

EP - 382

JO - Acta Orthopaedica

JF - Acta Orthopaedica

SN - 1745-3682

IS - 4

ER -