Measurements of knee rotation in vivo - Development and evaluation of an external device

Research output: ThesisDoctoral Thesis (compilation)

Abstract

The overall aim of this work was to evaluate a newly developed measurement device,
the Rottometer, for measuring knee rotation in different flexion angles with different
applied torques, in order to establish the normal range of healthy knee rotation reference
values and to study possible differences due to age and gender, as well as possible
differences between patients with habitual dislocating patella (HDP) and healthy
controls. The validity of the Rottometer was evaluated by simultaneous registrations
with roentgen stereometric analysis (RSA) (paper I). The two methods showed high
correlations concerning the total knee rotation at 90° and 60° of knee flexion angles
with 3, 6 and 9 Nm applied torques. The Rottometer was also concluded to be a reliable
measurement device concerning the one-week-apart and within day intra-tester
as well as the inter-tester reliability at 90°, 60° and 30° with 6 and 9 Nm as well as the
examiner’s apprehension of end-feel (paper II). In total, 120 knee healthy subjects (60
females and 60 males) equally distributed in four different age groups (15-30, 31-45,
46-60 and ≥ 61 years) were examined at 90°, 60° and 30° of knee flexion angles with
6 and 9 Nm applied torques as well as the examiner’s apprehension of end-feel (paper
III). No differences were found concerning the different flexion angles, between the left
and right knee or between the different age groups within the genders. However, the
females showed a 10-20 % larger range of knee rotation than the males at all different
flexion angles and applied torques. The knee rotation was also examined in 20 patients
(15 females and 5 males) with HDP (paper IV). No differences were found between
the affected and unaffected knees within the subjects. In accordance with the healthy
reference population, the female subjects showed a 10-20 % significantly larger range
of rotation than the male subjects in the HDP group, and no differences were found
between the patients affected with HDP and age matched healthy controls.

Details

Authors
  • Per Otto Almquist
Organisations
Research areas and keywords

Subject classification (UKÄ) – MANDATORY

  • Physiotherapy

Keywords

  • Knee, rotation, measurement, evaluation, gender, age, dislocating patella
Original languageEnglish
QualificationDoctor
Awarding Institution
Supervisors/Assistant supervisor
  • Charlotte Ekdahl, Supervisor
  • Thomas Fridén, Supervisor
Award date2012 Oct 26
Publisher
  • Department of Health Sciences, Lund University
Print ISBNs978-91-87189-17-3
Publication statusPublished - 2012
Publication categoryResearch

Bibliographic note

Defence details Date: 2012-10-26 Time: 09:00 Place: Health Sciences Centre, Hörsal 01 External reviewer(s) Name: Kvist, Joanna Title: docent Affiliation: Inst för medicin och hälsa, avd för sjukgymnastik, Linköpings Universitet --- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Division of Physiotherapy (Closed 2012) (013042000)

Total downloads

No data available

Related research output

Almquist, P. O., Ekdahl, C., Isberg, P-E. & Fridén, T., 2012, In : Journal of Orthopaedic Research.

Research output: Contribution to journalArticle

Almquist, P. O., Arnbjörnsson, A., Zätterström, R., Ryd, L., Ekdahl, C. & Fridén, T., 2002, In : Journal of Orthopaedic Research. 20, 3, p. 427-432

Research output: Contribution to journalArticle

View all (2)