Reference dosimetry at the neutron capture therapy facility at Studsvik

Research output: Contribution to journalArticle

Abstract

The purpose of this publication was to present and evaluate the methods for reference dosimetry in the epithermal neutron beam at the neutron capture therapy facility at Studsvik. Measurements were performed in a PMMA phantom and in air using ionization chambers and activation probes in order to calibrate the epithermal neutron beam. Appropriate beam-dependant calibration factors were determined using Monte Carlo methods for the detectors used in the present publication. Using the presented methodology, the. photon, neutron and total absorbed dose to PMMA was determined with an estimated uncertainty of +/-5.0%, +/-25%, and +/-5.5% (2 SD), respectively. The uncertainty of the determination of the photon absorbed dose was comparable to the case in conventional radiotherapy, while the uncertainty of the neutron absorbed dose is much higher using the present methods. The thermal neutron group fluence, i.e., the neutron fluence in the energy interval 0-0.414 eV, was determined with an estimated uncertainty of +/-2.8% (2 SD), which is acceptable for dosimetry in epithermal neutron beams. (C) 2003 American Association of Physicists in Medicine.