4D dosimetry and its applications to pre-treatment quality control and real-time in vivo dosimetry of VMAT treatments

Forskningsoutput: Kapitel i bok/rapport/Conference proceedingKonferenspaper i proceeding

Abstract

In this study, a 4D dosimetry concept was developed. This concept included a method for calculation of 3D reference absorbed dose matrices at every control point of the delivery using a clinical treatment planning system (TPS). Further, the gamma evaluation method was extended to incorporate the 4th dimension of the TPS calculated dose distributions. The applications of the 4D dosimetry concept on pre-treatment quality control and real-time in vivo dosimetry were investigated.

Detaljer

Författare
Enheter & grupper
Forskningsområden

Ämnesklassifikation (UKÄ) – OBLIGATORISK

  • Radiologi och bildbehandling
Originalspråkengelska
Titel på värdpublikation7th International Conference on 3D Radiation Dosimetry (IC3DDose)
FörlagIOP Publishing
Sidor012021
Volym444
StatusPublished - 2013
PublikationskategoriForskning
Peer review utfördJa
Evenemang7th International Conference on 3D Radiation Dosimetry (IC3DDose) - Sydney, Australien
Varaktighet: 2012 nov 42012 nov 8

Publikationsserier

Namn
Volym444
ISSN (tryckt)1742-6588
ISSN (elektroniskt)1742-6596

Konferens

Konferens7th International Conference on 3D Radiation Dosimetry (IC3DDose)
LandAustralien
OrtSydney
Period2012/11/042012/11/08