Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems: Failure Analysis and Computed Tomography Diagnostics in Two Cases.

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskrift

Abstract

We report on two patients with fracture of a modular, tapered and distally fixed, uncemented titanium revision hip stem, not previously described. A failure analysis revealed that the cause of the fractures was the development of fatigue cracks in the mid-stem cobalt-chromium modular junction ending in corrosion-fatigue failure. No material defects or stress risers were found in any of the implants. The diameter of the mid-stem modular junction might be undersized for use in heavy and active patients. We also report a new way of detecting an undisplaced fracture at the modular junction, using the scout image from a computed tomography (CT) scan; a technique that can be used when plain radiographs are inconclusive.

Detaljer

Författare
Enheter & grupper
Forskningsområden

Ämnesklassifikation (UKÄ) – OBLIGATORISK

  • Ortopedi
Originalspråkengelska
Sidor (från-till)850-855
TidskriftJournal of Arthroplasty
Volym29
Utgåva nummer4
StatusPublished - 2014
PublikationskategoriForskning
Peer review utfördJa