High temporal resolution monitoring of fermentations using an on-line amperometric flow-through microdetector

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskrift

Abstract

An on-line microdetector containing amperometric biosensors was used for high temporal resolution (t(90%) approximate to 120 s) monitoring of glucose and ethanol concentrations during small scale fermentations. ne ability of the microdetector to report on the effect of different experimental conditions was tested in fermentation processes carried out at 30 and at 37 degrees C. An increased ethanol production rate accompanied by an increased glucose consumption rate in the fermentation carried out at 37 degrees C was promptly revealed. Therefore, the microdetector proved to be an especially useful tool to monitor fermentations where the investigated processes are too fast to be followed by classical analytical approaches.

Detaljer

Författare
  • Kinga Zor
  • Szilveszter Gaspar
  • Masatoshi Hashimoto
  • Hiroaki Suzuki
  • Elisabeth Csöregi
Enheter & grupper
Forskningsområden

Ämnesklassifikation (UKÄ) – OBLIGATORISK

  • Analytisk kemi

Nyckelord

Originalspråkengelska
Sidor (från-till)43-48
TidskriftElectroanalysis
Volym19
Utgåva nummer1
StatusPublished - 2007
PublikationskategoriForskning
Peer review utfördJa