Strahlendosis bei renaler Vergrosserungsangiographie im Vergleich mit konventioneller Angiographie

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskrift

Abstract

Measurements of skin dose, integrated skin dose area, and integral dose in combined conventional and magnification renal angiography using lanthanoxibromide screens MR 600 and carbon fibre plate in filmchangertop have been made. The mean skin dose in magnification technique was 0.56 rad./film with focus-skin distance of 35 cm, FFD 1 m, 85-90 kV and sagittal diameter of abdomen of 20 cm. The combination of angiographic technique with and without magnification and the use of a high-sensitivity receiving unit makes it possible to keep the total skin dose per investigation (8.5 rad.) within the limits accepted in conventional abdominal angiography.

Detaljer

Författare
Enheter & grupper
Forskningsområden

Ämnesklassifikation (UKÄ) – OBLIGATORISK

  • Radiologi och bildbehandling
Bidragets titel på inmatningsspråkRadiation dose in renal magnification angiography compared with conventional angiography
Originalspråktyska
Sidor (från-till)175-178
TidskriftRadiologe
Volym18
Utgivningsnummer5
StatusPublished - 1978
PublikationskategoriForskning
Peer review utfördJa