Projekt per år
Sammanfattning
PURPOSE: The partial volume effect (PVE) is an important source of bias in brain perfusion measurements. The impact of tissue PVEs in perfusion measurements with dynamic susceptibility contrast MRI (DSC-MRI) has not yet been well established. The purpose of this study was to suggest a partial volume correction (PVC) approach for DSC-MRI and to study how PVC affects DSC-MRI perfusion results.
METHODS: A linear mixed perfusion model for DSC-MRI was derived and evaluated by way of simulations. Twenty healthy volunteers were scanned twice, including DSC-MRI, arterial spin labeling (ASL), and partial volume measurements. Two different algorithms for PVC were employed and assessed.
RESULTS: Simulations showed that the derived model had a tendency to overestimate perfusion values in voxels with high fractions of cerebrospinal fluid. PVC reduced the tissue volume dependence of DSC-MRI perfusion values from 44.4% to 4.2% in gray matter and from 55.3% to 14.2% in white matter. One PVC method significantly improved the voxel-wise repeatability, but PVC did not improve the spatial agreement between DSC-MRI and ASL perfusion maps.
CONCLUSION: Significant PVEs were found for DSC-MRI perfusion estimates, and PVC successfully reduced those effects. The findings suggest that PVC might be an important consideration for DSC-MRI applications. Magn Reson Med, 2016. © 2016 Wiley Periodicals, Inc.
Originalspråk | engelska |
---|---|
Sidor (från-till) | 2203–2214 |
Tidskrift | Magnetic Resonance in Medicine |
Volym | 77 |
Nummer | 6 |
Tidigt onlinedatum | 2016 juni 20 |
DOI | |
Status | Published - 2017 |
Ämnesklassifikation (UKÄ)
- Radiologi och bildbehandling
- Annan fysik
Fingeravtryck
Utforska forskningsämnen för ”A linear mixed perfusion model for tissue partial volume correction of perfusion estimates in dynamic susceptibility contrast MRI: Impact on absolute quantification, repeatability, and agreement with pseudo-continuous arterial spin labeling”. Tillsammans bildar de ett unikt fingeravtryck.-
-
MRI brain perfusion quantification at 3 tesla using arterial spin labeling
2006/07/02 → 2017/02/17
Projekt: Forskning