An attenuation method for reducing count rate losses in preclinical PET during intratherapeutic imaging

Emma Mellhammar, Magnus Dahlbom, Susan Evans-Axelsson, Johan Axelsson, Thuy Tran, Sven-Erik Strand

Forskningsoutput: Kapitel i bok/rapport/Conference proceedingKonferenspaper i proceedingPeer review

Sammanfattning

In pre-clinical imaging, tumor response to radionuclide therapy can be monitored with PET imaging. Radionuclides used for therapy such as 177Lu emit a significant amount of low energy photons. These photons may have an energy high enough to penetrate the imaged object and are the prone to be detected. Although these phtons are likely to be rejected electronically, they add dead-time to the system since they need to be processed by the electronics. This is a problem in high-sensitivity pre-clinical PET system with a low number of readout channels, such as the Genisys G8 investigated in this work. The low energy gammas may also affect image quality due to increased probability of pulse pile-up. The use of high-attenuating shields designed to absorb most of the low energy photons emitted from the therapeutic radionuclide were investigated. Cylindrical led shields were constructed with thicknesses between 1 and 3 mm. A 3mm thick cylindrical shield was also constructed out of Rose metal (50% Bi, 28% Pb, and 22% Sn). The diameter of the shield was wide enough to accommodate a NEMA IQ phantom and a mouse. The attenuation of the shields for annihilation radiation was measured with a 22Na point source placed at the center of the FOV. Measurements of the coincidence rate were performed with the lead shields in place. At a of thicknesses of 1 and 3mm, the coincidence rate was reduced by a factor or 0.70 and 0.40, respectively. To study the effect of the presence of a background of low energy gammas on the coincidence count rate and the efficacy of the lead shields, 177Lu was added to a 1 cm diameter hollow sphere. In the presence of 100 MBq of 177Lu, the coincidence count rate was reduced by a factor 0.20 due to the detector dead-time. Although the count rate was reduced by a factor of 0.40 with the 3mm shield around the source, the dead-time effects due to the 177Lu background were less than 7%. 18F imaging of a NEMA phantom and a tumor bearing mouse showed dramatic image distortions in the presence of the 177Lu background. When imaging of with the 3mm shield in place, the image distortions were eliminated and were comparable in to the images acquired without the background activity.

Originalspråkengelska
Titel på värdpublikation2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016
FörlagIEEE - Institute of Electrical and Electronics Engineers Inc.
Volym2017-January
ISBN (elektroniskt)9781509016426
DOI
StatusPublished - 2017 okt. 16
Evenemang2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016 - Strasbourg, Frankrike
Varaktighet: 2016 okt. 292016 nov. 6

Konferens

Konferens2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016
Land/TerritoriumFrankrike
OrtStrasbourg
Period2016/10/292016/11/06

Ämnesklassifikation (UKÄ)

  • Radiologi och bildbehandling

Fingeravtryck

Utforska forskningsämnen för ”An attenuation method for reducing count rate losses in preclinical PET during intratherapeutic imaging”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här