Sammanfattning
We present a case of combination of a leadless pacemaker (Micra) and a subcutaneous implantable cardioverter-defibrillator (S-ICD). The patient had a total of nine adequate shock treatments of ventricular fibrillation during 18 months of follow-up after the implantation. The shock treatments did not lead to any alteration in the Micra. All three sensing vectors of the S-ICD worked well. After 18 months, the functioning of both Micra and S-ICD continues to be uneventful. This case demonstrates that S-ICD combined with Micra may be a safe and feasible approach to provide pacing and ICD treatment without intracardiac leads.
Originalspråk | engelska |
---|---|
Sidor (från-till) | 1-3 |
Antal sidor | 3 |
Tidskrift | Journal of Electrocardiology |
Volym | 56 |
DOI | |
Status | Published - 2019 |
Ämnesklassifikation (UKÄ)
- Kardiologi
- Medicinsk apparatteknik