Determination of trace levels of dinitrophenolic compounds in environmental water samples using hollow fiber supported liquid membrane extraction and high performance liquid chromatography

Tarekegn Berhanu, Jing-fu Liu, Roberto Romero, N Megersa, Jan Åke Jönsson

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

A hollow fiber supported liquid membrane extraction method for the liquid chromatographic determination of dinitrophenolic compounds at ppt levels has been developed. Different variables affecting the extraction process, such as extraction time, shaking speed, acceptor pH, acceptor buffer concentration, salt content and humic acids have been studied. Enrichment factors up to 7000 times were obtained. Validation of the method included calibration experiments and studies of the linearity of the responses in different matrices. Good linearity was obtained in the environmental matrices evaluated. Detection limits range from 6.0 to 8.0 ng/L, and the relative standard deviations do not exceed 7% in terms of repeatability. (c) 2005 Elsevier B.V. All rights reserved.
Originalspråkengelska
Sidor (från-till)1-8
TidskriftJournal of Chromatography A
Volym1103
Nummer1
DOI
StatusPublished - 2006

Bibliografisk information

The information about affiliations in this record was updated in December 2015.
The record was previously connected to the following departments: Analytical Chemistry (S/LTH) (011001004)

Ämnesklassifikation (UKÄ)

  • Analytisk kemi

Fingeravtryck

Utforska forskningsämnen för ”Determination of trace levels of dinitrophenolic compounds in environmental water samples using hollow fiber supported liquid membrane extraction and high performance liquid chromatography”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här