Evaluation of 3D printed contrast detail phantoms for mammography quality assurance

Forskningsoutput: Kapitel i bok/rapport/Conference proceedingKonferenspaper i proceedingPeer review

Sammanfattning

Objects created by 3D printers are increasingly used in various medical applications. Today, affordable 3D printers, using Fused Deposition Modeling are widely available. In this project, a commercially available 3D printer was used to replicate a conventional radiographic contrast detail phantom. Printing materials were selected by comparing their x-ray attenuation properties. Two replicas were printed using polylactic acid, with different filling patterns. The printed phantoms were imaged by a clinical mammography system, using automatic exposure control. Phantom images were visually and quantitively compared to images of the corresponding conventional contrast detail phantom. Visual scoring of the contrast detail elements was performed by a medical physics student. Contrast-to-noise ratio (CNR) was calculated for each phantom element. The diameter and thickness of the smallest visible phantom object were 0.44 mm and 0.09 mm, respectively, for both filling patterns. For the conventional phantom, the diameter and thickness of the smallest visible object were 0.31 mm and 0.09 mm. Visual inspection of printed phantoms revealed some linear artefacts. These artefacts were however not visible on mammographic projections. Quantitively, average CNR of printed phantom objects followed the same trend with an increase of average CNR with increasing disk height. However, there is a limitation of detail objects with disk diameters below 1.25 mm, caused by the available nozzle size. Based upon the encouraging results, future work will explore the use of different materials and smaller nozzle diameters.

Originalspråkengelska
Titel på värdpublikation16th International Workshop on Breast Imaging, IWBI 2022
RedaktörerHilde Bosmans, Nicholas Marshall, Chantal Van Ongeval
FörlagSPIE
ISBN (elektroniskt)9781510655843
DOI
StatusPublished - 2022
Evenemang16th International Workshop on Breast Imaging, IWBI 2022 - Leuven, Belgien
Varaktighet: 2022 maj 222022 maj 25

Publikationsserier

NamnProceedings of SPIE - The International Society for Optical Engineering
Volym12286
ISSN (tryckt)0277-786X
ISSN (elektroniskt)1996-756X

Konferens

Konferens16th International Workshop on Breast Imaging, IWBI 2022
Land/TerritoriumBelgien
OrtLeuven
Period2022/05/222022/05/25

Ämnesklassifikation (UKÄ)

  • Medicinsk bildbehandling
  • Radiologi och bildbehandling

Fingeravtryck

Utforska forskningsämnen för ”Evaluation of 3D printed contrast detail phantoms for mammography quality assurance”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här