Sammanfattning
Tactile displays have a wide range of applications in assistive technology, prosthetics, and rehabilitation. Vibrotactile stimulation shows potential to be integrated into wearable devices because of its small size, low power consumption, and relative fast responding speed. In this paper, experiments investigating vibrotactile minimal detectable level, just noticeable difference, and two point discrimination distance using eccentric rotating mass (ERM) and linear resonant actuator (LRA) are described and the results were analyzed. LRAs have shown potential to transmit binary information, while ERMs are more suitable for conveying more complexed signals. For both LRAs and ERMs, it is difficult to for subjects to correctly distinguish two points within 45 mm.
Originalspråk | engelska |
---|---|
Titel på värdpublikation | 2016 6th IEEE International Conference on Biomedical Robotics and Biomechatronics, BioRob 2016 |
Förlag | IEEE - Institute of Electrical and Electronics Engineers Inc. |
Sidor | 1266-1271 |
Antal sidor | 6 |
ISBN (elektroniskt) | 9781509032877 |
ISBN (tryckt) | 9781509032884 |
DOI | |
Status | Published - 2016 juli 26 |
Evenemang | 6th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics, BioRob 2016 - Singapore, Singapore Varaktighet: 2016 juni 26 → 2016 juni 29 |
Konferens
Konferens | 6th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics, BioRob 2016 |
---|---|
Land/Territorium | Singapore |
Ort | Singapore |
Period | 2016/06/26 → 2016/06/29 |
Ämnesklassifikation (UKÄ)
- Robotik och automation