Ferrocene-mediated thermal biosensor

Bin Xie, Masoud Khayyami, Titus Nwosu, Per-Olof Larsson, Bengt Danielsson

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

A biosensor that combines electrochemistry and calorimetry was investigated. Using this sensor, glucose measurements were performed by employing glucose oxidase together with ferrocene as electron-transfer mediator. The electrochemical reaction was accomplished by applying a voltage between a platinum column (working electrode), in contact with a crushed reticulated vitreous carbon (RVC) matrix onto which glucose oxidase was immobilized, and platinum wires (counter electrodes) located at the inlet and outlet of the column. For detection, the thermal signal generated by the redox reaction was measured as opposed to measuring the electrochemical signal. By using this method, a linear range of glucose concentration up to 20 mmol l-1 was achieved.

Originalspråkengelska
Sidor (från-till)845-848
Antal sidor4
TidskriftThe Analyst
Volym118
Nummer7
DOI
StatusPublished - 1993

Ämnesklassifikation (UKÄ)

  • Analytisk kemi

Fingeravtryck

Utforska forskningsämnen för ”Ferrocene-mediated thermal biosensor”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här