INVESTIGATION OF PARTIAL VOLUME EFFECT IN DIFFERENT PET/CT SYSTEMS: A COMPARISON OF RESULTS USING THE MADEIRA PHANTOM AND THE NEMA NU-2 2001 PHANTOM.

L Chipiga, Marie Sydoff, I Zvonova, Christian Bernhardsson

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

2 Citeringar (SciVal)

Sammanfattning

Positron emission tomography combined with computed tomography (PET/CT) is a quantitative technique used for diagnosing various diseases and for monitoring treatment response for different types of tumours. However, the accuracy of the data is limited by the spatial resolution of the system. In addition, the so-called partial volume effect (PVE) causes a blurring of image structures, which in turn may cause an underestimation of activity of a structure with high-activity content. In this study, a new phantom, MADEIRA (Minimising Activity and Dose with Enhanced Image quality by Radiopharmaceutical Administrations) for activity quantification in PET and single photon emission computed tomography (SPECT) was used to investigate the influence on the PVE by lesion size and tumour-to-background activity concentration ratio (TBR) in four different PET/CT systems. These measurements were compared with data from measurements with the NEMA NU-2 2001 phantom. The results with the MADEIRA phantom showed that the activity concentration (AC) values were closest to the true values at low ratios of TBR (<10) and reduced to 50 % of the actual AC values at high TBR (30-35). For all scanners, recovery of true values became closer to 1 with an increasing diameter of the lesion. The MADEIRA phantom showed good agreement with the results obtained from measurements with the NEMA NU-2 2001 phantom but allows for a wider range of possibilities in measuring image quality parameters.
Originalspråkengelska
Sidor (från-till)365-70
TidskriftRadiation Protection Dosimetry
Tidigt onlinedatum2016 mar 2
DOI
StatusPublished - 2016 jun

Ämnesklassifikation (UKÄ)

  • Radiologi och bildbehandling

Fingeravtryck

Utforska forskningsämnen för ”INVESTIGATION OF PARTIAL VOLUME EFFECT IN DIFFERENT PET/CT SYSTEMS: A COMPARISON OF RESULTS USING THE MADEIRA PHANTOM AND THE NEMA NU-2 2001 PHANTOM.”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här