Measurement of the electrically evoked compound action potential via a neural response telemetry system

N Dillier, WK Lai, Bengt Almqvist, C Frohne, J Muller-Deile, M Stecker, E von Wallenberg

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

The main aim of this study was to validate a new technique, neural response telemetry (NRT), for measuring the electrically evoked compound action potential in adult cochlear implant users via their Nucleus CI24M implant. Thirty-eight adults were evaluated with a variety of measurement procedures with the NRT software. Electrically evoked compound action potentials were obtained in 31 of the 38 adults (81.6%) and in 132 of the 160 electrodes (82.5%) tested. In addition to validating this technique, We also established a set of default clinical test parameters.
Originalspråkengelska
Sidor (från-till)407-414
TidskriftAnnals of Otology, Rhinology & Laryngology
Volym111
Nummer5
StatusPublished - 2002