T1 - Monte Carlo investigation of PET [68Ga]Ga-DOTA-TOC activity-administration protocols for consistent image quality
AU - Kalaitzidis, Philip
AU - Gustafsson, Johan
AU - Hindorf, Cecilia
AU - Ljungberg, Michael
PY - 2023/9
Y1 - 2023/9
N2 - One example of a PET exam that suffers from noise problems is [68Ga]Ga-DOTA-TOC, where patients are generally administered between 100 and 200 MBq [68Ga]Ga-DOTA-TOC, irrespective of size. However, a fixed activity can result in low signal-to-noise ratios (SNRs) in larger patients. This study aimed to evaluate the impact on image quality with respect to injected activity and patient habitus through Monte Carlo (MC) simulation. Eight anthropomorphic computer phantoms with body mass indices (BMIs) between 19 kg/m2 and 38 kg/m2 and tumours distributed in the liver were simulated using the MC software Gate v8.2 with an activity distribution defined according to [68Ga]Ga-DOTA-TOC standardised uptake values. Three activity-administration protocols were simulated: (i) with a fixed activity of 100 MBq, (ii) with the activity scaled by 2 MBq/kg, and (iii) with the activity scaled by a body size-dependent power-function based on the SNR obtained with (ii). BMI, weight, body surface area, and abdominal circumference were evaluated body size parameters. Images were reconstructed with the CASToR software and evaluated for background SNR and lesion contrast-to-noise ratio (CNR). Large SNR variabilities were obtained with protocols (i) and (ii), while (iii) generated good consistency. Several tumours failed to reach a CNR of 5 for large phantoms with protocol (i), but the CNR was generally improved by (ii) and (iii). An activity scaled by patient habitus generate better image quality consistency, which increases the likelihood that patients receive a similar standard of care.
AB - One example of a PET exam that suffers from noise problems is [68Ga]Ga-DOTA-TOC, where patients are generally administered between 100 and 200 MBq [68Ga]Ga-DOTA-TOC, irrespective of size. However, a fixed activity can result in low signal-to-noise ratios (SNRs) in larger patients. This study aimed to evaluate the impact on image quality with respect to injected activity and patient habitus through Monte Carlo (MC) simulation. Eight anthropomorphic computer phantoms with body mass indices (BMIs) between 19 kg/m2 and 38 kg/m2 and tumours distributed in the liver were simulated using the MC software Gate v8.2 with an activity distribution defined according to [68Ga]Ga-DOTA-TOC standardised uptake values. Three activity-administration protocols were simulated: (i) with a fixed activity of 100 MBq, (ii) with the activity scaled by 2 MBq/kg, and (iii) with the activity scaled by a body size-dependent power-function based on the SNR obtained with (ii). BMI, weight, body surface area, and abdominal circumference were evaluated body size parameters. Images were reconstructed with the CASToR software and evaluated for background SNR and lesion contrast-to-noise ratio (CNR). Large SNR variabilities were obtained with protocols (i) and (ii), while (iii) generated good consistency. Several tumours failed to reach a CNR of 5 for large phantoms with protocol (i), but the CNR was generally improved by (ii) and (iii). An activity scaled by patient habitus generate better image quality consistency, which increases the likelihood that patients receive a similar standard of care.
KW - Ga]Ga-DOTA-TOC
KW - Anthropomorphic computer phantoms
KW - Monte Carlo simulation
KW - Positron emission tomography
UR - https://www.scopus.com/pages/publications/85169511011
U2 - 10.1016/j.heliyon.2023.e19504
DO - 10.1016/j.heliyon.2023.e19504
M3 - Article
C2 - 37681181
AN - SCOPUS:85169511011
SN - 2405-8440
VL - 9
JO - Heliyon
JF - Heliyon
IS - 9
M1 - e19504
ER -