No gadolinium K‐edge detected on the first clinical photon‐counting computed tomography scanner

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

Purpose
This study aimed to elucidate whether gadolinium contrast in clinically relevant doses can be used with photon-counting computed tomography (PCCT) as an alternative contrast agent in clinical applications.

Material/methods
A CTDI phantom with 3D printed rods filled with different concentrations of gadolinium and iodine contrast was scanned in a PCCT and an energy-integrated computed tomography (EICT). Attenuation values at different monoenergetic steps were extracted for each contrast concentration.

Results
For PCCT, gadolinium reached an attenuation >100 HU (103 HU) at 40 keV with a concentration 5 mmol/L whereas the same level was reached at 50 keV (118 HU) for 10 mmol/L and 90 keV (114 HU) for 25 mmol/L. For iodine, the same level of attenuation was reached at 100 keV (106 HU) with a concentration 8.75 mg I/mL.

For EICT the lowest gadolinium contrast concentration needed to reach >100 HU (108 HU) was 10 mmol/L at 50 keV. For 25 mmol/L 100 HU was reached at 100 keV. For iodine contrast 108 HU was reached at 110 keV for 8.75 mg I/mL.

Conclusion
No K-edge potential or difference in attenuation curves between iodine and gadolinium contrast is detected on the first clinical available PCCT. Clinically relevant attenuation levels were barely achieved in this setting with gadolinium concentrations approved for human use. The results of this study suggest that, given current scanning technology, gadolinium is not a clinically useful contrast agent for computed tomography because no K-edge was detected.
Originalspråkengelska
Artikelnummere14324
Sidor (från-till)1-7
TidskriftJournal of Applied Clinical Medical Physics
DOI
StatusE-pub ahead of print - 2024 mars 12

Ämnesklassifikation (UKÄ)

  • Radiologi och bildbehandling

Fingeravtryck

Utforska forskningsämnen för ”No gadolinium K‐edge detected on the first clinical photon‐counting computed tomography scanner”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här