Optimization of energy and directional response of a small 4π silicon dosemeter for quality control of CT/CBCT-units – A 3D CAD, Monte-Carlo, AM approach

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

A 4π radiation dosemeter for use in medical radiology was designed. It is based on a solid state silicon detector, a sensor wafer, a flex card, a 3D plastic holder and a spherical stainless steel filter with a distribution of holes around the detector. The detector is attached to the wafer using only low Z material. To achieve an energy and directional response which is as uniform as possible for various radiation qualities and beam directions, the filter was designed using a series of Monte Carlo calculations. The energy filter and its hole pattern were created using Additive Manufacturing (AM) in the form of metal 3D printing. The functionality of the dosemeter was designed to fulfill the quality criteria of a combined energy and angular dependence of less than 5% for the IEC beam qualities RQR and RQT in the range of 65–145 kV. This is a major improvement to the existing solutions that may need software corrections to be used for the same beam quality range.

Originalspråkengelska
Artikelnummer109050
TidskriftApplied Radiation and Isotopes
Volym158
DOI
StatusPublished - 2020

Ämnesklassifikation (UKÄ)

  • Radiologi och bildbehandling

Fingeravtryck

Utforska forskningsämnen för ”Optimization of energy and directional response of a small 4π silicon dosemeter for quality control of CT/CBCT-units – A 3D CAD, Monte-Carlo, AM approach”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här