Radiative MRI coil design using parasitic scatterers: MRI Yagi

Juan-Diego Sanchez, Johan Avendal, Adnan Bibic, Buon Kiong Lau

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

322 Nedladdningar (Pure)

Sammanfattning

Conventionally, radiofrequency (RF) coils used for magnetic resonance imaging (MRI) are electrically small and designed for nearfield operation. Therefore, existing antenna design techniques are mostly irrelevant for RF coils. However, the use of higher frequencies in ultrahigh field (UHF) MRI allows for antenna design techniques to be adapted to RF coil designs. This study proposes the use of parasitic scatterers to improve the performance of an existing 7T MRI coil called the single-sided adapted dipole (SSAD) antenna. The results reveal that scatterers arranged in a Yagi fashion can be applied to reduce local specific absorption rate (SAR) maxima of a reference SSAD by 40% with only a 6% decrease in the propagated B1+ field at the tissue depth of 15 cm. The higher directivity of the proposed design also decreasing the coupling with additional elements, making this antenna suitable for use in high density arrays. These findings show the potential of parasitic scatterers as an effective method to improve the performance of existing radiative MRI coils.
Originalspråkengelska
Sidor (från-till)1570-1575
Antal sidor6
TidskriftIEEE Transactions on Antennas and Propagation
Volym66
Nummer3
DOI
StatusPublished - 2018 mars

Ämnesklassifikation (UKÄ)

  • Annan elektroteknik och elektronik

Fingeravtryck

Utforska forskningsämnen för ”Radiative MRI coil design using parasitic scatterers: MRI Yagi”. Tillsammans bildar de ett unikt fingeravtryck.