Sammanfattning
A sensitive and selective amperometric biosensor, based on NADH-oxidase (NADOx) working in tandem with horseradish peroxidase (HRP), which is electrically connected to graphite electrode via Os-redox polymer (RP), was developed. Two designs were considered for the biosensor optimization: (i) the first one, [G/HRP-NADOx], exploited the direct electron transfer between HRP and the graphite electrode (G); (ii) the second one, [G/RP-HR-P-NADOx], used R-P as mediator. In both cases the enzyme matrix was cross-linked using poly(ethylenglycol) diglycidyl ether. The biosensors sensitivities, estimated as the slopes of linear ranges from the calibration curves, recorded in a single line flow injection setup, showed that G/RP-HRP-NADOx electrode (1.62 +/- 0.05 mA M-1) presents a higher efficiency than G/HRP-NADOx electrode (0.30 +/- 0.03 mA M-1).
| Originalspråk | engelska |
|---|---|
| Sidor (från-till) | 755-760 |
| Tidskrift | Revue Roumaine de Chimie |
| Volym | 54 |
| Nummer | 9 |
| Status | Published - 2009 |
Bibliografisk information
The information about affiliations in this record was updated in December 2015.The record was previously connected to the following departments: Biotechnology (LTH) (011001037), Analytical Chemistry (S/LTH) (011001004)
Ämnesklassifikation (UKÄ)
- Analytisk kemi
- Industriell bioteknik