Sammanfattning
A capillary filling microfluidic proteomic sample processing system has been realized in an automated set-up. Chip-integrated solid-phase extraction is performed in a 96-array format followed by sequential capillary action elution into a piezoelectric microdispenser and subsequent transfer to MALDI targets. Samples are eluted and deposited in volumes of 200–300 nl. The robotic system offers calibration to user-defined microextraction arrays and MALDI-target formats at micrometre resolution. Built-in force feedback control ensures a precise and robust microchip docking/handling in three dimensions. An efficient automated washing protocol eliminates analyte carry-over. System throughput ranges typically from 50–100 samples h-1.
Originalspråk | engelska |
---|---|
Sidor (från-till) | 3147-3153 |
Tidskrift | Measurement Science & Technology |
Volym | 17 |
Nummer | 12 |
DOI | |
Status | Published - 2006 |
Bibliografisk information
The information about affiliations in this record was updated in December 2015.The record was previously connected to the following departments: Biomedical Engineering (011200011), Robotics (ceased) (LUR000026), Analytical Chemistry (S/LTH) (011001004), Departments at LTH (011200000)
Ämnesklassifikation (UKÄ)
- Analytisk kemi
- Medicinteknik