The upgraded photon tagging facility at the MAX IV Laboratory

Jan-Olof Adler, Mark Boland, Jason Brudvik, Kevin Fissum, Kurt Hansen, Lennart Isaksson, Per Lilja, Lars-Johan Lindgren, Magnus Lundin, Björn Nilsson, Dmytro Pugachov, Anders Sandell, Bent Schröder, Vladimir Avdeichikov, Pavel Golubev, Bo Jakobsson, J. R. M. Annand, K. Livingston, R. Igarashi, L. MyersA. Nathan, W. J. Briscoe, G. Feldman, M. Kovash, D. Branford, K. Foehl, P. Grabmayr, V. Takau, G. O'Rielly, D. Burdeynyi, V. Ganenko, V. Morochovskyi, G. Vashchenko

Forskningsoutput: TidskriftsbidragArtikel i vetenskaplig tidskriftPeer review

Sammanfattning

A description is given of the upgraded photon tagging facility at the MAX IV Laboratory. Two magnetic spectrometers are used to momentum analyze post-bremsstrahlung electrons. The tagged photon range extends from 10 to 180 MeV with an energy resolution of about 300 keV. The system has been operated at rates up to 4 x 10(6) photons s(-1) MeV (-1). Different diagnostic tools are described as well as the experimental program.

Ämnesklassifikation (UKÄ)

  • Acceleratorfysik och instrumentering

Fingeravtryck

Utforska forskningsämnen för ”The upgraded photon tagging facility at the MAX IV Laboratory”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här