Two-dimensional acoustic focusing of microparticles in two-phase droplet-based microfluidic systems increases particle detectability

M. Ohlin, A. Fornell, M. Tenje

Forskningsoutput: Kapitel i bok/rapport/Conference proceedingKonferenspaper i proceedingPeer review

Sammanfattning

We have fabricated a silicon-glass two-phase droplet-based microfluidic system and implemented two-dimensional acoustic focusing prior to droplet generation as well as continuously throughout the whole system to increase particle detectability. Using acoustic focusing we have effectively minimized sedimentation of the encapsulated particles and thereby increased particle detectability by as much as 44% compared to unactuated operation of the system.

Originalspråkengelska
Titel på värdpublikation20th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2016
FörlagChemical and Biological Microsystems Society
Sidor713-714
Antal sidor2
ISBN (elektroniskt)9780979806490
StatusPublished - 2016
Evenemang20th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2016 - Dublin, Irland
Varaktighet: 2016 okt. 92016 okt. 13

Konferens

Konferens20th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2016
Land/TerritoriumIrland
OrtDublin
Period2016/10/092016/10/13

Ämnesklassifikation (UKÄ)

  • Medicinsk laboratorie- och mätteknik

Fingeravtryck

Utforska forskningsämnen för ”Two-dimensional acoustic focusing of microparticles in two-phase droplet-based microfluidic systems increases particle detectability”. Tillsammans bildar de ett unikt fingeravtryck.

Citera det här